Patient Radiation Dose Received from the Body Interventional Radiology

Authors

  • Somsak Wongsanon
  • Wichai Witchathorntakun
  • Wattana Wongsanon

Abstract

Objectives and Principles: To make a risk assessment and find out the amount of radiation dose which patients received from the Body Interventional Radiology that it will be safe or not when compared with reference standard dose.

Methods : A descriptive historical research is done by using radiation units of Srinakarin Hospital Diagnostic Radiology Department, Faculty of Medicine, Khon-Kaen University and the data receive from112 patients. Started from January-June 2007 and then were analyzed the relationship between the sample radiation dose and reference radiation dose.

วัตถุประสงค์ : 1. เพื่อศึกษาปริมาณรังสีที่ผู้ป่วยได้รับจากการตรวจทางรังสีร่วมรักษาระบบลำตัวว่าแต่ละหัตถการมีค่าเท่าใด Result : Itsfound that the abdominal Embolization. Takes the longest fluoroscopic time of about 15.7 minutes and the patients received the highest accumulated dose of 133.73 Gy.cm2 in mean average and the maximum radiation dose for each patient was 0.78 Gy. TransArterialchemoembolization(TOCE) เป็นหัตถการที่มีความถี่ในการตรวจมากที่สุดใช้เวลา fluoroscopic time เฉลี่ย 8.58 นาทีโดยผู้ป่วยได้รับปริมาณรังสีสะสมเฉลี่ย 121.09 Gy.CM 2 และผู้ป่วยได้รับปริมาณรังสีสูงสุดต่อรายที่ 327.60 Gy.CM 2 และ Percutaneus Tranhepatic Biliary Drainage ( PTBD)เป็นหัตถการที่ Chemo-embolization procedure which tookmost of the fluoroscopic time at approximately 8.58 minutes and the average accumulated dose was 121.09 Gy.cm2. The maximum accumulated dose in a person was 0.73 Gy. Percutaneus transhepatic biliary drainage was the procedureใช้เวลา fluoroscopic time เฉลี่ยน้อยที่สุด that took less of fluoroscopic time about4.71นาที   โดยผู้ป่วยจะได้รับปริมาณรังสีสะสมเฉลี่ย 24.127 Gy.CM 2 และผู้ป่วยได้รับปริมาณรังสีสูงสุดต่อรายที่ 115.89 Gy.CM 2 และเมื่อนำค่าปริมาณรังสีที่ได้ไปวิเคราะห์ความสัมพันธ์กับระดับรังสีที่ถือว่าปลอดภัยโดยICRP 4.71 minutes. The patients received the average accumulated dose was about 24.127 Gy.cm 2 and the maximum dose was about 0.34 Gy in a person. All the dose were then analyzed and compared with the standard reference dose. It was found that in every procedure the entrance skin dose were 0.090, 0.008 and 0.002 Gy. Respectively and the rests were less than safety standard value of 2 Gy.

Conclusion : Patients receive accumulated skin dose less than 2 Gy which is conformed to  the standard dose limit. It indicates that the process of body interventional radiology is absolutely safe.

Downloads

How to Cite

1.
Wongsanon S, Witchathorntakun W, Wongsanon W. Patient Radiation Dose Received from the Body Interventional Radiology. SRIMEDJ [Internet]. 2013 Nov. 1 [cited 2024 Dec. 27];24(4):339-43. Available from: https://li01.tci-thaijo.org/index.php/SRIMEDJ/article/view/13044

Issue

Section

Original Articles