Comparison of coaxial circle circuit with carbon dioxide absorber and coaxial D circuit without carbon dioxide absorber in general anesthesia.

Authors

  • Wattana Tantanatewin
  • Ammara Leesan
  • Sunchai Theerapongpakdee

Abstract

Abstract

This  clinical  study  is  to  compare  the  use  of  coaxial  circle  circuit  with  CO2  absorber  and  coaxial  D  circuit  without   CO2  absorber  in  20  patients  of  physical  status  ASA  class  I  and  II. The  mean  age,  body  weight,  and  hemoglobin  concentration  are  40 ± 9.94  years  old,  56.77 ± 7.33 kgs.,  11.76 ± 1.22  gms/dl  respectively.  Fresh  gas  flow  (FGF)  and  arterial  blood  gas  was  taken  after  stabilizing  period  of  15 – 30 minutes.  Coxial  D  circuit  without  CO2 absorber  was  used  secondly  and  arterial  blood  gas  was  taken  in  the  same  manner.  The  pH,  PaCO2,  PaO2  values  were  compared  by  paired  T-test  statistical   analysis.  It  is  concluded  that  at  6  lpm.  Of  FGF  and  minute  ventilation,  there  is  statistically  significant  difference  (P <0.05)  for  pH  value.  However  the values  of  pH, PaCO2  and  PaO2  are  still  within  normal  limit  in  both  circuits.

Downloads

How to Cite

1.
Tantanatewin W, Leesan A, Theerapongpakdee S. Comparison of coaxial circle circuit with carbon dioxide absorber and coaxial D circuit without carbon dioxide absorber in general anesthesia. SRIMEDJ [Internet]. 2013 Nov. 20 [cited 2024 Nov. 24];11(3):107-12. Available from: https://li01.tci-thaijo.org/index.php/SRIMEDJ/article/view/13881

Issue

Section

Original Articles