Simple Circuit for Continuous Monitoring of Mean Arterial Blood Pressure.
บทคัดย่อ
A comparative study of the three blood pressure monitoring devices were carried out. The mean arterial pressure (MAP) recorded by a pressure transducer were not significant difference (P>0.05) when compared to the value recorded by a simple device. There were significant difference (P>0.05) in the MAP recorded by the pressure transducer compared to the normal auscultatory blood pressure measurement (NABP). The MAP recorded by the simple device were also not significant difference (P>0.05) when compared to the NABP.
The simple device is economical and can be used for the continuous monitoring of the mean arterial blood pressure in critical patient in provincial hospital.ดาวน์โหลด
รูปแบบการอ้างอิง
1.
Prasertchuang C, Tienthong S, Tantanatawin W, Pengsakes N, Vorpratum S, Tantisawee P. Simple Circuit for Continuous Monitoring of Mean Arterial Blood Pressure. SRIMEDJ [อินเทอร์เน็ต]. 11 พฤศจิกายน 2013 [อ้างถึง 28 ธันวาคม 2025];1(1):7-10. available at: https://li01.tci-thaijo.org/index.php/SRIMEDJ/article/view/13235