Comparison of AflaTest-P and Afla-M1 Column for Aflatoxin M1 Analysis in Milk
Keywords:
aflatoxin M1, fluorometer, HPLC, milkAbstract
The amount of aflatoxin M1 (AFM1) in milk using AflaTest-P column and Afla-M1 column was compared. Fluorometer and high performance liquid chromatography (HPLC) were used as detectors for both kinds of column. It was found that HPLC was appropriate for both types, resulting in a 78% recovery of 1 ppb spiked milk from AflaTest-P column and 86% for Afla-M1 column. However, using AflaTest-M as a calibration standard, fluorometer was proven not to be appropriate for neither types of column.