A Headspace Solid Phase Microextraction Method for Using to Monitor Hexanal and Heptanal Content in Food Samples
Keywords:
hexanal, heptanal, headspace solid-phase microextractionAbstract
Headspace solid phase microextraction (HS-SPME) technique coupled with Gas Chromatography (GC) was developed to measure hexanal and heptanal (off-flavour compounds, occurred from lipid oxidation) in some food samples. The HS-SPME and GC parameters were studied. Six food
samples; infant cereal, instant noodle, potato French fried, deep fried peanut, fried rice (Khawtan) and fried pork skin (Capmoo) were analyzed. Hexanal and heptanal were found in the range of 0.15-10.40 μg/g and 0.14-1.65 μg/g, respectively. Hexanal content in Capmoo increased significantly during storage. It was found that increase of lipid content in food samples can reduce the analysis recoveries.
Downloads
Published
How to Cite
Issue
Section
License
online 2452-316X print 2468-1458/Copyright © 2022. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/),
production and hosting by Kasetsart University of Research and Development Institute on behalf of Kasetsart University.